Diagnostic Accuracy of the Quidel Sofia Rapid Influenza Fluorescent Immunoassay in Patients with Influenza-like Illness: A Systematic Review and Meta-analysis

Article information

Tuberc Respir Dis. 2021;84(3):226-236
Publication date (electronic) : 2021 May 13
doi : https://doi.org/10.4046/trd.2021.0033
1Department of Internal Medicine, Jeju National University Hospital, Jeju National University School of Medicine, Jeju, Republic of Korea
2Division of Pulmonary and Critical Care Medicine, Department of Internal Medicine, Kangbuk Samsung Hospital, Sungkyunkwan University School of Medicine, Seoul, Republic of Korea
3Department of Pulmonary and Critical Care Medicine, Kyung Hee University Hospital at Gangdong, Kyung Hee University School of Medicine, Seoul, Republic of Korea
Address for correspondence: Jonghoo Lee, M.D., Ph.D., Department of Internal Medicine, Jeju National University Hospital, Jeju National University School of Medicine, Aran 13 gil 15, Jeju 63241, Republic of Korea, Phone: 82-64-717-1601, Fax: 82-64-717-1131, E-mail: lovlet@jejunu.ac.kr
*These authors contributed equally to this work.
Received 2021 February 24; Revised 2021 April 4; Accepted 2021 May 3.

Abstract

Background

Although the Quidel Sofia rapid influenza fluorescent immunoassay (FIA) is widely used to identify influenza A and B, the diagnostic accuracy of this test remains unclear. Thus, the objective of this study was to determine the diagnostic performance of this test compared to reverse transcriptase-polymerase chain reaction.

Methods

A systematic literature search was performed using MEDLINE, EMBASE, and the Cochrane Central Register. Pooled sensitivity, specificity, diagnostic odds ratio (DOR), and a hierarchical summary receiver-operating characteristic curve (HSROC) of this test for identifying influenza A and B were determined using meta-analysis. A sensitivity subgroup analysis was performed to identify potential sources of heterogeneity within selected studies.

Results

We identified 17 studies involving 8,334 patients. Pooled sensitivity, specificity, and DOR of the Quidel Sofia rapid influenza FIA for identifying influenza A were 0.78 (95% confidence interval [CI], 0.71–0.83), 0.99 (95% CI, 0.98–0.99), and 251.26 (95% CI, 139.39–452.89), respectively. Pooled sensitivity, specificity, and DOR of this test for identifying influenza B were 0.72 (95% CI, 0.60–0.82), 0.98 (95% CI, 0.96–0.99), and 140.20 (95% CI, 55.92–351.54), respectively. The area under the HSROC for this test for identifying influenza A was similar to that for identifying influenza B. Age was considered a probable source of heterogeneity.

Conclusion

Pooled sensitivities of the Quidel Sofia rapid influenza FIA for identifying influenza A and B did not quite meet the target level (≥80%). Thus, caution is needed when interpreting data of this study due to substantial between-study heterogeneity.

Introduction

Influenza, an acute respiratory viral infection caused by influenza A or B virus, occurs mainly in the winter months throughout the world. It causes significant morbidity and mortality worldwide1,2. Adequate antiviral therapy can shorten the time of illness and reduce the duration of hospitalization and the risk of complications from influenza infections3. Clinical benefit of antiviral therapy is the greatest when it is started soon after the onset of influenza illness4. Therefore, rapid and accurate diagnosis of influenza infection is necessary in clinical practice.

Although polymerase chain reaction (PCR) has been used as the reference standard for diagnosing viral infections, performing PCR is relatively expensive. In addition, it requires technical expertise5. Alternatively, point-of-care rapid influenza diagnostic tests (RIDTs) can detect viral antigens by immunoassay and provide quick results within 30 minutes. They can facilitate antiviral therapy, reduce additional diagnostic tests and hospitalization therapy, and induce appropriate infection control measures6,7. A recent systematic review and meta-analysis for 162 diagnostic accuracy studies of RIDTs has revealed that traditional RIDTs have specificities higher than 98% with poor sensitivities (54.4% for influenza A and 53.2% for influenza B)5. Two classes of RIDTs, automated immunochromatographic antigen detection tests (digital immunoassays [DIAs]) and rapid nucleic acid amplification tests (NAATs), have been used since 20115. Pooled sensitivities for DIAs (80.0% for influenza A and 76.8% for influenza B) and rapid NAATs (91.6% for influenza A and 95.4% for influenza B) are significantly higher than those for traditional RIDTs5.

As a type of DIA, the Quidel Sofia rapid influenza fluorescent immunoassay (FIA) (Quidel Corp., San Diego, CA, USA) is a point-of-care test to detect influenza A and B in less than 15 minutes using a compact instrument (Sofia analyzer). Although the performance characteristics of this test to detect seasonal influenza virus strains have been established, the diagnostic accuracy of the Quidel Sofia rapid influenza FIA is not yet fully known. Thus, the objective of this systematic review and meta-analysis of clinical trial data was to investigate diagnostic properties of the Quidel Sofia rapid influenza FIA in patients with influenza like illness.

Materials and Methods

1. Data sources and search strategy

This meta-analysis is reported in accordance with the Preferred Reporting Items for Systematic reviews and Meta-Analyses of Diagnostic Test Accuracy Studies statement8. A comprehensive search of three electronic databases (MEDLINE, EMBASE, and the Cochrane Central Register) up to July of 2020 was performed. Search terms for influenza included “Influenza, Human” [MeSh] OR “Influenza A virus” OR “Influenza B virus” OR “influenza” OR “flu”. Search terms for the tests included “rapid test*” OR “rapid diagnos*” OR “point-of-care test*” OR “immunoassay*” OR “immunochromatographic test*” OR “influenza FIA” OR “Quidel Sofia Influenza” OR “Rapid Detection Flu”. As this study was a systematic review of published articles, neither informed consent nor ethics approval was required. A manual search of references listed in relevant review articles was also conducted.

2. Study selection

We included studies that met the following inclusion criteria: (1) full-length reports published in peer-reviewed English language journals; (2) studies that evaluated the performance of the Quidel Sofia rapid influenza FIA, compared to a reference standard; (3) studies that included patients with influenza-like illness; and (4) studies that provided sufficient data to calculate absolute numbers of true-positive, false-positive, false-negative, and true-negative results. Review articles, case reports, commentaries, and studies reporting outcomes without raw data or peer review were excluded. Demographics and underlying diseases of participants were not restricted.

Influenza-like illness was defined as fever ≥38°C and any signs/symptoms of respiratory tract infection (e.g., cough, sputum, sore throat, wheezing, etc.). We allowed the followings as specimens: nasopharyngeal aspirates, swabs, or washes; nasal aspirates, swabs, or washes; saliva; and throat swabs. A reference standard was either a commercial reverse transcriptase–polymerase chain reaction (RT-PCR) or a laboratory-developed RT-PCR.

3. Data extraction and quality assessment

Two authors independently reviewed potentially relevant studies and each study according to predefined eligibility criteria, after which data were extracted. Any disagreements that arose during the process of study selection or data extraction were resolved by discussion. A predefined form was used to extract data from each study. The following data from each study included in the meta-analysis were extracted: author, year of publication, study design, place of study, number of participants, age, proportion of children, sex, study period, type of reference standard, type of specimens, and type of population. As recommended by the Cochrane Collaboration, the Quality Assessment of Diagnostic Accuracy Studies (QUADAS)-2 tool was used to assess the risk of bias in diagnostic test accuracy9. Discrepancies were resolved by consensus between the two authors.

4. Data synthesis and statistical analysis

For diagnostic meta-analysis, random effects meta-analyses were performed to generate pooled estimates with 95% confidence intervals (CIs). Numbers of patients were extracted with true-positive, false-positive, false-negative, and true negative test results either directly or indirectly through a recalculation based on reported measures of accuracy in combination with the prevalence and sample size of the included study. Pooled sensitivity and specificity, positive likelihood ratio (PLR), negative likelihood ratio (NLR), diagnostic odds ratio (DOR), and area under the receiver-operating characteristic curve (AUC) were calculated as pooled estimates with a 95% CI10. Hierarchical summary receiver-operating characteristic curves (HSROCs) were also constructed. The heterogeneity was evaluated using the Higgins I2 statistics on a scale of 0–1. An I2>0.5 indicated a substantial level of between-study heterogeneity. To assess effects of potential sources of heterogeneity, subgroup analyses were performed using the following covariates to the model: study design (single vs. multicenter), number of participants (≥250 vs. <250), study period (influenza vs. non-specific season), and study population (children vs. adult). Pooled sensitivity and specificity estimates were calculated for each covariate. To investigate the effect of study quality, sensitivity analyses were performed. A p-value <0.05 was considered statistically significant. All statistical analyses were performed with Stata statistical software version 14.2 (Stata Corp LP, College Station, TX, USA) and Review Manager version 5.3 (Nordic Cochrane Centre, The Cochrane Collaboration, Copenhagen, Denmark).

Results

1. Study search, characteristics of included studies, and quality of included studies

The literature search process is shown in Figure 1. Initially, 432 articles from PubMed, 1,310 articles from EMBASE, 356 articles from the Cochrane library, and an additional article from hand-searching were identified. After removing duplicate articles, 1,664 potentially eligible articles were screened. After reviewing titles and abstracts, 1,629 search records were removed and the remaining 35 articles were eligible for full text reading. Fifteen articles were excluded for reasons shown in Figure 1. With quantitative synthesis, seventeen studies were included in our final analysis1127.

Figure 1

Flow diagram showing the identification of eligible studies. RT-PCR: reverse transcription polymerase chain reaction.

Features of included studies are summarized in Table 1. For influenza A, we identified 17 studies involving 8,334 participants. For influenza B, sixteen studies involving 7,909 subjects met the defined inclusion criteria. One study assessed influenza A infection only21. The number of patients in each trial ranged from 68 to 1,649. All studies were published between 2012 and 2018. Most studies evaluated combined populations of adults and children.

Characteristics of studies included in the meta-analysis

Results of the QUADAS-2 assessment are shown in Figure 2. For patient selection, the index test, and the reference standard domain, more than 50% of included studies were judged to have a low risk of bias. However, for the study flow and the timing domain, the risk of bias was high or unclear for 52.9% of the included studies because of unclear intervals between the index test and the reference standard. Considering our inclusion criteria, we had little concern for the applicability of results from selected studies for each domain.

Figure 2

(A) Summary of risk of bias in included studies. (B) Risk of bias graph for included studies.

2. Diagnostic accuracy of the Quidel Sofia rapid influenza FIA to identify influenza A and B

Figures 3 and 4 show paired forest plots of sensitivity and specificity of the Quidel Sofia rapid influenza FIA for detecting influenza A and B. The pooled sensitivity across studies of the Quidel Sofia rapid influenza FIA for identifying influenza A was 0.78 (95% CI, 0.71–0.83), with a pooled specificity of 0.99 (95% CI, 0.98–0.99). The pooled PLR and NLR were 56.99 (95% CI, 31.87–101.90) and 0.23 (95% CI, 0.18–0.29), respectively. The DOR for influenza A was 251.26 (95% CI, 139.39–452.89).

Figure 3

Paired forest plots of sensitivity and specificity of the Quidel Sofia rapid influenza fluorescent immunoassay for detecting influenza A. TP: true positive; FP: false positive; FN: false negative; TN: true negative; CI: confidence interval.

Figure 4

Paired forest plots of sensitivity and specificity of the Quidel Sofia rapid influenza fluorescent immunoassay for detecting influenza B. TP: true positive; FP: false positive; FN: false negative; TN: true negative; CI: confidence interval.

The pooled sensitivity across studies for identifying influenza B was 0.72 (95% CI, 0.60–0.82), with a pooled specificity of 0.98 (95% CI, 0.96–0.99). The pooled PLR and NLR were 40.08 (95% CI, 17.26–93.07) and 0.29 (95% CI, 0.19–0.42), respectively. The DOR for influenza B was 140.20 (95% CI, 55.92–351.54). The pooled sensitivity and specificity of the Quidel Sofia rapid influenza FIA for identifying influenza A and B were not significantly different (p=0.341 for sensitivity and p=0.206 for specificity). Figure 5 shows HSROCs for the index test. AUCs of the Quidel Sofia rapid influenza FIA for identifying influenza A and influenza B were similar (0.96 with 95% CI of 0.94–0.98 for influenza A vs. 0.95 with 95% CI of 0.92–0.96 for influenza B, p=0.166).

Figure 5

Hierarchical summary receiver operating characteristic curves of the Quidel Sofia rapid influenza fluorescent immunoassay for detecting influenza A (A) and influenza B (B). HSROC: hierarchical summary receiver-operating characteristic curve; AUC: area under the receiver-operating characteristic curve.

3. Subgroup and sensitivity analyses for the Quidel Sofia rapid influenza FIA to identify influenza A and B

The Higgins I2 statistics proved significant heterogeneity for both the sensitivity (0.91 with 95% CI of 0.88–0.94 for influenza A and 0.87 with 95% CI of 0.82–0.92 for influenza B) (Figure 3) and specificity (0.92 with 95% CI of 0.90–0.95 for influenza A and 0.98 with 95% CI of 0.97–0.98 for influenza B) (Figure 4). Subgroup analyses were also performed to investigate potential sources of heterogeneity (Table 2). The sensitivity of the Quidel Sofia rapid influenza FIA was significantly increased when tests were performed in children (0.86 with 95% CI of 0.78–0.92 for influenza A and 0.79 with 95% CI of 0.71–0.85 for influenza B) than when tests were performed in adults (0.74 with 95% CI of 0.67–0.79 for influenza A and 0.33 with 95% CI of 0.10–0.65 for influenza B).

Subgroup analysis for the diagnostic performance of the Quidel Sofia rapid influenza fluorescent immunoassay

In sensitivity analysis to investigate the influence of each individual study on the overall analysis estimate, one study had a significantly different sensitivity than other studies on influenza A24. Even after exclusion of that study24, the pooled sensitivity across studies on influenza A was similar to that of overall studies (0.79 with 95% CI of 0.75–0.82). Instead, the heterogeneity decreased (0.51 with 95% CI of 0.23–0.79).

Discussion

According to the rule set by the Food and Drug Administration (FDA), RIDTs for influenza A and B are required to have a sensitivity of at least 80% and a specificity of at least 95% compared to an FDA-cleared nucleic acid based-test or other currently appropriate tests and FDA accepted comparator methods other than a correctly performed viral culture method28. A recent systematic review and meta-analysis performed a search up to May 2017 and compared accuracies of traditional RIDTs, rapid NAATs, and DIAs in patients with suspected influenza5. For diagnosis of influenza A and B, pooled sensitivities of DIAs and the Quidel Sofia rapid influenza FIA were 80.0% and 76.8%, respectively5. In the present study, compared to RT-PCR, the pooled sensitivity of the Quidel Sofia rapid influenza FIA to identify influenza A and B were 78% and 72%, respectively. Our findings did not quite reach the target level of sensitivity required by the FDA. Therefore, some patients with negative results on the Quidel Sofia rapid influenza FIA might still need be confirmed to have an influenza infection by an alternative diagnostic method that is more sensitive.

Influenza type could affect the accuracy of RIDTs. A previous meta-analysis has revealed that overall RIDTs show an increased sensitivity for detecting influenza A than for detecting influenza B (64.6% vs. 52.2%; p=0.05)29. Influenza A virus can cause more severe disease, higher influenza-associated hospitalization, and more death than influenza B virus29. Higher virulence of influenza A might lead to higher viral burden, which can result in a relatively higher sensitivity19. In the present study, although the pooled sensitivity of the Quidel Sofia rapid influenza FIA for identifying influenza A tended to be higher than that for identifying influenza B, the difference between the two was not statistically significant.

The type of specimen might also lead to difference in diagnostic performances. Although we tried to perform subgroup analyses according to specimens, we could only find two suitable studies26,27. One study compared the results of testing throat and NP swabs specimens26. The sensitivity for each type of specimens was 72% and 100% for detecting influenza A and 71% and 100% for detecting influenza B, respectively26. In the other study, sensitivities of the Quidel Sofia rapid influenza FIA with saliva specimens were comparable to those with NP swabs specimens27. The sensitivity of the test with NP swabs samples was significantly higher than that with saliva samples for detecting influenza A (74.2% vs. 59.2%, p=0.014) and similar between two samples in influenza B (75.9% vs. 65.5%, p=0.387)27. Based on results of these two studies regarding different specimens used for detection, the sensitivity of the Quidel Sofia rapid influenza might be lower for detecting influenza A and B viruses using oropharynx or saliva samples.

The Quidel Sofia rapid influenza FIA has advantages of being simple, fast, and easy for viral testing. The pooled specificity of this tool in our study was approximately 98%, above the target level for detecting both influenza A and B. From these findings, we believe that clinicians could diagnose influenza with assurance on the basis of a positive result from the Quidel Sofia rapid influenza FIA.

Large heterogeneities are commonly reported for systematic reviews of studies on diagnostic test accuracy30. Substantial between-study heterogeneity among enrolled studies was also observed in the present study. Age is a probable source of heterogeneity for between-studies in pooled estimates. In the present study, pooled sensitivities of the Quidel Sofia rapid influenza FIA were significantly higher in children (by approximately 12% higher for influenza A and 46% higher for influenza B) than in adults. The duration of influenza virus shedding is commonly measured from the time of symptom onset to the time of shedding cessation. Children have been reported to have a tendency to shed the virus for a longer duration than adults31. Longer duration of influenza virus shedding in children might be associated with a higher sensitivity of this test in children than in adults. However, because the number of studies that distinguished children from adults was very small, our findings should be interpreted with caution.

The sensitivity observed in one study24 was significantly lower than that observed in most studies. In that study, an older patient population (median age of 57 years) and a study protocol that did not specify the need for particular symptoms or duration of illness might have contributed to the significantly lower sensitivity24. These factors might have been affected by low virus shedding24. In our sensitivity analysis conducted after excluding this study24, the between-study heterogeneity was decreased.

To the best of our knowledge, this is the first meta-analysis to investigate the Quidel Sofia rapid influenza FIA for detecting influenza. However, potential limitations of the present study should be considered when interpreting our results. First, because this present study was based on a relatively small number of trials, our results should be carefully interpreted due to its limited statistical power. Second, we could not make an assessment for publication bias since no reliable methods existed to investigate this issue for diagnostic test accuracy studies32, Finally, as a sample for viral diagnosis, nasopharyngeal aspirates could show higher quality than nasopharyngeal swabs. As mentioned previously, although we tried to investigate the diagnostic accuracy of the Quidel Sofia rapid influenza FIA according to the type of samples, we were unable to perform such analysis because of data limitations.

In conclusion, we found that pooled sensitivities of the Quidel Sofia rapid influenza FIA were slightly below the target level set by the FDA for both influenza A and B. Therefore, physicians should consider the possibility of false-negative results by this test, especially for adults. Although pooled specificities of this test was very high for both influenza A and B, substantial between-study heterogeneity requires careful interpretation of the data.

Notes

Authors’ Contributions

Conceptualization: Lee J. Methodology: Lee J, Song JU. Formal analysis: Lee J, Song JU. Data curation: Lee J, Song JU. Investigation: Lee J, Song JU. Writing - original draft preparation: Lee J, Song JU. Writing - review and editing: Lee J, Kim YH. Approval of final manuscript: all authors.

Conflicts of Interest

No potential conflict of interest relevant to this article was reported.

Funding

This work was supported by a research grant from the Jeju National University Hospital Research Fund of Jeju National University in 2020.

References

1. World Health Organization. Use of influenza rapid diagnostic tests [Internet] Geneva: World Health Organization; 2010. [cited 2020 Oct 12]. Available from: https://www.who.int/influenza/resources/publications/influenza_rapid_test/en/ .
2. Chow EJ, Rolfes MA, O’Halloran A, Alden NB, Anderson EJ, Bennett NM, et al. Respiratory and nonrespiratory diagnoses associated with influenza in hospitalized adults. JAMA Netw Open 2020;3:e201323.
3. Atkins CY, Patel A, Taylor TH Jr, Biggerstaff M, Merlin TL, Dulin SM, et al. Estimating effect of antiviral drug use during pandemic (H1N1) 2009 outbreak, United States. Emerg Infect Dis 2011;17:1591–8.
4. Fry AM, Goswami D, Nahar K, Sharmin AT, Rahman M, Gubareva L, et al. Efficacy of oseltamivir treatment started within 5 days of symptom onset to reduce influenza illness duration and virus shedding in an urban setting in Bangladesh: a randomised placebo-controlled trial. Lancet Infect Dis 2014;14:109–18.
5. Merckx J, Wali R, Schiller I, Caya C, Gore GC, Chartrand C, et al. Diagnostic accuracy of novel and traditional rapid tests for influenza infection compared with reverse transcriptase polymerase chain reaction: a systematic review and meta-analysis. Ann Intern Med 2017;167:394–409.
6. Benito-Fernandez J, Vazquez-Ronco MA, Morteruel-Aizkuren E, Mintegui-Raso S, Sanchez-Etxaniz J, Fernandez-Landaluce A. Impact of rapid viral testing for influenza A and B viruses on management of febrile infants without signs of focal infection. Pediatr Infect Dis J 2006;25:1153–7.
7. Bonner AB, Monroe KW, Talley LI, Klasner AE, Kimberlin DW. Impact of the rapid diagnosis of influenza on physician decision-making and patient management in the pediatric emergency department: results of a randomized, prospective, controlled trial. Pediatrics 2003;112:363–7.
8. McInnes MD, Moher D, Thombs BD, McGrath TA, Bossuyt PM, et al. PRISMA-DTA Group. Preferred reporting items for a systematic review and meta-analysis of diagnostic test accuracy studies: the PRISMA-DTA statement. JAMA 2018;319:388–96.
9. Whiting P, Rutjes AW, Reitsma JB, Bossuyt PM, Kleijnen J. The development of QUADAS: a tool for the quality assessment of studies of diagnostic accuracy included in systematic reviews. BMC Med Res Methodol 2003;3:25.
10. Reitsma JB, Glas AS, Rutjes AW, Scholten RJ, Bossuyt PM, Zwinderman AH. Bivariate analysis of sensitivity and specificity produces informative summary measures in diagnostic reviews. J Clin Epidemiol 2005;58:982–90.
11. Bruning AH, van Dijk K, van Eijk HW, Koen G, van Woensel JB, Kruisinga FH, et al. Evaluation of a rapid antigen detection point-of-care test for respiratory syncytial virus and influenza in a pediatric hospitalized population in the Netherlands. Diagn Microbiol Infect Dis 2014;80:292–3.
12. Busson L, Mahadeb B, De Foor M, Vandenberg O, Hallin M. Contribution of a rapid influenza diagnostic test to manage hospitalized patients with suspected influenza. Diagn Microbiol Infect Dis 2017;87:238–42.
13. Dunn J, Obuekwe J, Baun T, Rogers J, Patel T, Snow L. Prompt detection of influenza A and B viruses using the BD Veritor System Flu A+B, Quidel(R) Sofia(R) Influenza A+B FIA, and Alere BinaxNOW(R) Influenza A&B compared to real-time reverse transcription-polymerase chain reaction (RT-PCR). Diagn Microbiol Infect Dis 2014;79:10–3.
14. Gomez S, Prieto C, Folgueira L. A prospective study to assess the diagnostic performance of the Sofia((R)) Immunoassay for Influenza and RSV detection. J Clin Virol 2016;77:1–4.
15. Hazelton B, Gray T, Ho J, Ratnamohan VM, Dwyer DE, Kok J. Detection of influenza A and B with the Alere i Influenza A & B: a novel isothermal nucleic acid amplification assay. Influenza Other Respir Viruses 2015;9:151–4.
16. Hazelton B, Nedeljkovic G, Ratnamohan VM, Dwyer DE, Kok J. Evaluation of the Sofia Influenza A + B fluorescent immunoassay for the rapid diagnosis of influenza A and B. J Med Virol 2015;87:35–8.
17. Kammerer PE, Radin JM, Hawksworth AW, Myers CA, Brice GT. Performance of the Quidel Sofia rapid influenza diagnostic test during the 2012–2013 and 2013–2014 influenza seasons. Influenza Other Respir Viruses 2016;10:220–3.
18. Lee CK, Cho CH, Woo MK, Nyeck AE, Lim CS, Kim WJ. Evaluation of Sofia fluorescent immunoassay analyzer for influenza A/B virus. J Clin Virol 2012;55:239–43.
19. Leonardi GP, Wilson AM, Zuretti AR. Comparison of conventional lateral-flow assays and a new fluorescent immunoassay to detect influenza viruses. J Virol Methods 2013;189:379–82.
20. Lewandrowski K, Tamerius J, Menegus M, Olivo PD, Lollar R, Lee-Lewandrowski E. Detection of influenza A and B viruses with the Sofia analyzer: a novel, rapid immunofluorescence-based in vitro diagnostic device. Am J Clin Pathol 2013;139:684–9.
21. Noh JY, Choi WS, Lee J, Kim HL, Song JY, Cheong HJ, et al. Clinical performance of the Sofia Influenza A+B FIA in adult patients with influenza-like illness. Diagn Microbiol Infect Dis 2015;83:130–2.
22. Ryu SW, Lee JH, Kim J, Jang MA, Nam JH, Byoun MS, et al. Comparison of two new generation influenza rapid diagnostic tests with instrument-based digital readout systems for influenza virus detection. Br J Biomed Sci 2016;73:115–20.
23. Ryu SW, Suh IB, Ryu SM, Shin KS, Kim HS, Kim J, et al. Comparison of three rapid influenza diagnostic tests with digital readout systems and one conventional rapid influenza diagnostic test. J Clin Lab Anal 2018;32:e22234.
24. Selove W, Rao LV. Performance of rapid SOFIA Influenza A+B test compared to Luminex x-TAG respiratory viral panel assay in the diagnosis of influenza A, B, and subtype H3. J Investig Med 2016;64:905–7.
25. Tuttle R, Weick A, Schwarz WS, Chen X, Obermeier P, Seeber L, et al. Evaluation of novel second-generation RSV and influenza rapid tests at the point of care. Diagn Microbiol Infect Dis 2015;81:171–6.
26. Yang JH, Huang PY, Shie SS, Yang S, Tsao KC, Wu TL, et al. Diagnostic performance of the Sofia(R) influenza A+B fluorescent immunoassay in adult outpatients in Northern Taiwan. J Med Virol 2018;90:1010–8.
27. Yoon J, Yun SG, Nam J, Choi SH, Lim CS. The use of saliva specimens for detection of influenza A and B viruses by rapid influenza diagnostic tests. J Virol Methods 2017;243:15–9.
28. Federal Register. Microbiology devices: reclassification of influenza virus antigen detection test systems intended for use directly with clinical specimens [Internet] Washington, DC: Federal Register; 2017. [cited 2020 Oct 12]. Available from: https://www.federalregister.gov/documents/2017/01/12/2017-00199/microbiology-devices-reclassification-of-influenza-virus-antigen-detection-test-systems-intended-for .
29. Chartrand C, Leeflang MM, Minion J, Brewer T, Pai M. Accuracy of rapid influenza diagnostic tests: a meta-analysis. Ann Intern Med 2012;156:500–11.
30. Dinnes J, Deeks J, Kirby J, Roderick P. A methodological review of how heterogeneity has been examined in systematic reviews of diagnostic test accuracy. Health Technol Assess 2005;9:1–113.
31. Ng S, Lopez R, Kuan G, Gresh L, Balmaseda A, Harris E, et al. The timeline of influenza virus shedding in children and adults in a household transmission study of influenza in Managua, Nicaragua. Pediatr Infect Dis J 2016;35:583–6.
32. Leeflang MM. Systematic reviews and meta-analyses of diagnostic test accuracy. Clin Microbiol Infect 2014;20:105–13.

Article information Continued

Figure 1

Flow diagram showing the identification of eligible studies. RT-PCR: reverse transcription polymerase chain reaction.

Figure 2

(A) Summary of risk of bias in included studies. (B) Risk of bias graph for included studies.

Figure 3

Paired forest plots of sensitivity and specificity of the Quidel Sofia rapid influenza fluorescent immunoassay for detecting influenza A. TP: true positive; FP: false positive; FN: false negative; TN: true negative; CI: confidence interval.

Figure 4

Paired forest plots of sensitivity and specificity of the Quidel Sofia rapid influenza fluorescent immunoassay for detecting influenza B. TP: true positive; FP: false positive; FN: false negative; TN: true negative; CI: confidence interval.

Figure 5

Hierarchical summary receiver operating characteristic curves of the Quidel Sofia rapid influenza fluorescent immunoassay for detecting influenza A (A) and influenza B (B). HSROC: hierarchical summary receiver-operating characteristic curve; AUC: area under the receiver-operating characteristic curve.

Table 1

Characteristics of studies included in the meta-analysis

Study Design Country No. of participants Age (yr) (mean or median) Proportion of children (%) Male (%) Study period Reference standard Type of specimens Population
Bruning11, 2014 Single center study Netherlands 68 NA 100 NA December 2013 to February 2014 RT-PCR NP swabs or aspirates Children (aged 0–16 years old) with symptoms of respiratory illness admitted to either pediatric ICU or infant ward
Busson12, 2017 Multi-center study Belgium 267 3.5 66.8 46.6 January 2015 to March 2015 RT-PCR NP swabs, NP aspirates, broncho-alveolar lavages Children and adult patients admitted with suspected influenza
Dunn13, 2014 Single center study USA 240 NA NA NA January 2013 to April 2013 RT-PCR Nasal wash Symptomatic patients under 18 years of age
Gomez14, 2016 Single center study Spain 1,065 NA NA NA November 2013 to April 2014 RT-PCR NP swabs, NP aspirates Adult (80.6%) and pediatric (19.4%) patients with respiratory tract symptoms hospitalized
Hazelton15, 2015 Single center study Australia 202 56 0 NA NA RT-PCR NP swabs Patients ≥16 yr with an influenza-like illness
Hazelton16, 2015 Single center study Australia 209 56 0 NA NA RT-PCR Throat swabs, NP swabs, or nose swabs Patients with an influenza-like illness
Kammerer17, 2016 Multi-center study USA 871 NA 90.0 NA 2012 to 2014 RT-PCR Nasal swabs Patients with an influenza-like illness (>70% of patients were under the age of 25 years)
Lee18, 2012 Single center study South Korea 169 27.7 NA 56.4 December 2011 to February 2012 RT-PCR and virus culture NP swabs Patients with an influenza-like illness
Leonardi19, 2013 Single center study USA 141 NA 48.0 NA Influenza seasons from 2006 to 2011 and the 2011–2012 RT-PCR NP swabs Frozen original influenza-positive specimens and prospective specimens
Lewandrowski20, 2013 Multi-center study USA 2,047 NA 93.1 52.9 NA RT-PCR or virus culture Nasal swabs and NP swabs or aspirates Patients with an influenza-like illness
Noh21, 2015 Multi-center study South Korea 391 40 NA 37.6 December 2012 to April 2013 RT-PCR NP swabs Adult patients with influenza-like illness
Ryu22, 2016 Single center study South Korea 314 30.4 NA 51.9 January 2014 to February 2015 RT-PCR NP swabs Patients showing influenza-like symptoms
Ryu23, 2018 Single center study South Korea 158 23.8 58.7 NA 2016 RT-PCR NP swabs Patients with an influenza-like illness between neonates and 90 years old
Selove24, 2016 Single center study USA 1,649 57 15.3 50 September 2014 to May 2015 RT-PCR Nasal aspirates Patients with an influenza-like illness
Tuttle25, 2015 Single center study Germany 686 3.5 for mean, 1.8 for median 100 55 December 2012 to April 2013 RT-PCR or virus culture NP or nasal swabs Patients aged 0–18 years with an influenza-like illness
Yang26, 2018 Single center study Taiwan 109 38.8 0 56.9 January 2012 to December 2013 RT-PCR and/or virus culture NP or throat swabs Patients who presented at out-patient clinics or the emergency department with influenza-like illness
Yoon27, 2017 Single center study South Korea 385 46 NA 46.5 December 2014 to April 2015 RT-PCR NP swabs or saliva Patients with an influenza-like illness

NA: not available; RT-PCR: reverse transcription polymerase chain reaction; ICU: intensive care unit.

Table 2

Subgroup analysis for the diagnostic performance of the Quidel Sofia rapid influenza fluorescent immunoassay

Variable No. of studies No. of patients Sensitivity Specificity


Adjusted (95% CI) p-value Adjusted (95% CI) p-value
Studies that evaluated influenza A

 Study design

  Single center 13 5,344 0.79 (0.72–0.85) 0.583 0.99 (0.98–1.00) 0.574

  Multicenter 4 2,990 0.75 (0.63–0.87) 0.99 (0.97–1.00)

 No. of participants

  ≥250 9 7,038 0.72 (0.65–0.79) 0.173 0.99 (0.98–1.00) 0.114

  <250 8 1,296 0.84 (0.77–0.90) 0.97 (0.95–0.99)

 Study period

  Influenza season 11 3,784 0.79 (0.73–0.86) 0.440 0.99 (0.98–1.00) 0.610

  Non-specific season 6 4,550 0.74 (0.64–0.85) 0.99 (0.97–1.00)

 Population

  Children 3 951 0.86 (0.78–0.92) 0.019 0.97 (0.96–0.98) 0.332

  Adults 2 593 0.74 (0.67–0.79) 0.97 (0.94–0.98)

Studies that evaluated influenza B

 Study design

  Single center 13 5,310 0.72 (0.59–0.84) 0.937 0.98 (0.97–1.00) 0.228

  Multicenter 3 2,599 0.73 (0.48–0.97) 0.97 (0.93–1.00)

 No. of participants

  ≥250 8 6,690 0.69 (0.53–0.84) 0.730 0.99 (0.97–1.00) 0.432

  <250 8 1,219 0.75 (0.60–0.90) 0.97 (0.94–1.00)

 Study period

  Influenza season 10 3,372 0.76 (0.63–0.89) 0.403 0.98 (0.96–1.00) >0.999

  Non-specific season 6 4,537 0.65 (0.45–0.85) 0.98 (0.96–1.00)

 Population

  Children 3 994 0.79 (0.71–0.85) <0.001 0.92 (0.90–0.94) 0.005

  Adults 1 202 0.33 (0.10–0.65) 0.99 (0.97–1.00)

CI: confidence interval.